Standard Electric Furnace (FO300)

Standard Electric Furnace (FO300)

Description

8 types of internal capacity are available, with program operation function

Features of Standard Electric Furnace (FO300)

Features
Able to do program operation
Features
Easier up/down key setting, equipped with the controller ofgood operation performance to realize the high precision of temp. control.
Features
Digital representation of setting temp., time and indicator value.
Features
Use the controller of digital set and overheat proection, in the event of abnormality it has safety controls such as self diagnosis and overcurrent ELB.
Features
The temp. sensor applies R-thermocouple with long service life.
Features
Equipped with exhaust port.
Features
It also has optional functions like air exhaust device, sample tray, N2 introduction device (attached with flowmeter), temp. output terminal, etc..

Additional information

Model

FO300

Temp. setting/display method

Digital setting by▲/▼ keys / Digital display

Weight

37kg

Item Code

214117

Operating temp. range

100 to 1,150℃

Temp. control accuracy

±2.0℃(at 1,150℃)

Max. temp. reaching time

Approx. 70min.

Exterior material

Cold rolled steel plate with baked-on melamine resin finish

Internal material

Ceramic fiber

Sensor

R-thermocouple

Heater

Iron-chrome wire

Heater Power

2kW

Exhaust port

18 mm I.D. (upper part)

Cooling Fan type

Axial fan motor

Temp. controller

PID control by microprocessor

Overheat protection

Integrated controller

Overheat protection set

Digital setting

Operation function

Fixed temp., program (6 modes 30 seg.×1, 15 seg.×2, 10 seg.×3)

Additional Function

Deviation correction, Power outage compensation, Key lock

Timer

1min to 99hr59min and 999hr 50min digital setting, Auto start, Quick auto stop

Safe Device

Self diagnosis circuit (Abnormal temp. sensor, Auto overheat protection), Averheat protector, Avercurrent ELB, key lock

Accessories

Exhaust port 1 pc.

Optional Stand

ON30C

Optional Others 2

Exhaust device (AC220V), Sample tray, N2 introduction device (Attached with flow meter), Recorder, Indicator lamp (Stand-by / running / Malfunction),

Optional Others

External communication (RS485), Temp. output terminal (4-20mA), Output terminal for alarm device, Time up output terminal

Internal dimensions(W×D×H mm)

200×250×150mm

External dimensions(W×D×H mm)

446×505×567mm

Remarks(dimensions)

7.5

Power Source

AC115/220V
Single-phase with step down transformer