Double Chamber Incubator (Natural Convection & Low Temp.) (INC820)

Double Chamber Incubator (Natural Convection & Low Temp.) (INC820)

Description

High-temp. and Low-temp. incubations can be carried out in each chamber individually in 1 unit

Features of Double Chamber Incubator (Natural Convection & Low Temp.) (INC820)

Feature
The lower chamber is equipped with cycle/manual defrost function.
Feature
Equipped with self-diagnostic function, key-lock, ELB for overcurrent, overheat protector
Feature
Each chamber is equipped with door lock key (2pcs. each)

Additional information

Model

INC820

Operation function (Upper Chamber)

Fixed temp., Quick auto stop, Auto stop, Auto start operation

Operating temp. range(Upper Chamber)

Room temp.+5℃~80℃

Temp. adjustment accuracy (Upper Chamber)

±0.5℃(at 37℃)

Temp. distribution accuracy (Upper Chamber)

±1.0℃(at 37℃)

Interior

Stainless Steel

Cable port (Upper Chamber)

30mm ID (Right side from the front)

Temp. adjustment accuracy (Lower Chamber)

±0.3℃(at continuous refrigeration)

Weight

160kg

Exhaust port (Upper Chamber)

30mm(with Damper), 2 ports on the top

Internal Dimensions (Upper Chamber)

600×530×500mm

Internal capacity (Upper Chamber)

150L

Shelf Step/ Pitch (Upper Chamber)

13 steps / 30mm

Standard Shelf / Support (Upper Chamber)

2 shelves/4 suppots

System (Lower Chmber)

Forced convection

Operating temp. range(Lower Chamber)

+4℃~50℃

Item Code

211210

System (Upper Chamber)

Natural convection, Air jacket

Controller (Upper Chamber)

VS3 Type, Fixed operation temperature controller

Shelf / support

3 shelves/6 pcs.

Refrigerator (Lower Chamber)

Air cooling, 250W

Cable port (Lower Chamber)

50mm ID (Right side from the front)

Controller (Lower Chamber)

VS4 Type Program operation temp. controller

Operation function (Lower Chamber)

Program operation(30 steps ×1、15 steps×2、10 steps×3), Fixed temp. operation, Quick auto stop, Auto stop, Auto start

Safety device (Lower Chamber)

Self-diagnostic functions(Temp. sensor error, Heater disconnection, SSR short-circuit, Main relay error, Automatic overheat prevention), Key-lock, Over current ELB, Overheat preventive device

Interior (Lower Chamber)

600×477×500mm

Shelf support step number/pitch

13 steps / 30mm

Temp. distribution accuracy (Lower Chamber)

±1.0℃(at continuous refrigeration)

Sensor

Pt100Ω

Internal capacity

143L

Shelf loading Capacity

Approx. 15kg/ shelf

Door Key

2 pcs. / each chamber

External dimensions(W×D×H mm)

710×656×1792mm

Remarks(dimensions)

150L (Upper chamber ) + 143L (Lower chamber)

Power Source

AC220V 7A
Single-phase with step-down transformer

Refrigerant (Lower Chamber)

R404A

Defrost Operation (Lower Chamber)

Manual / Cycle