Home » Products » Fume Hood (Energy Saving Structure) (LDN-N150S)
Energy Saving Structure (Push-pull ventilator type)
Features of Fume Hood (Energy Saving Structure) (LDN-N150S)
LDN-N150S
A003046
350kg
Cold-rolled steel plate (chemical resistance baking finish)
Non-asbestos and non-combustible decorative board (Baffle plate: multi slit system)
Ceramitite (S)
Tempered glass 6mm Balance weight
Single type 1pc
AC115/ 220V with dual grounding wire (dropout prevention) 1pc.
LED lamp x 1pc
Supply air from the front air wheel on the work surface (intake of indoor air)
Exhaust delay timer, Door falling prevention stopper, Flow monitor
Flow monitor (LED bar display)
9m3/min (*0.25m/sec) / 14m3/min (*0.4m/sec) *Front average wind speed
14Pa (0.25m/sec) / 34Pa (0.4m/sec)
1500×750×2400mm
115/220V V